5 Care, cleaning and maintenance by the practice team
5.4 Vacuum system
224
Refilling the cleaning agent
When the adjacent display appears in the status bar of the touchscreen,
the cleaning agent for cleaning the suction system is almost depleted. It
should then be refilled as soon as possible.
Vacuum system, Mixing with DENTOSEPT P
CAUTION
The cleaning agent for the vacuum system could be mistaken for the
treatment solution for the water lines.
➢ Do not fill DENTOSEPT P waterline treatment solution into the
cleaning agent tank for chemical suction hose cleaning! Use only
"Agents for the suction lines" approved by Sirona, see Care,
cleaning and disinfecting agents" [ → 201].
1. Open the maintenance flap on the base of the water unit. The
cleaning agent tank for chemical suction hose cleaning is located on
the left side.
2. Pull the tank slightly out of the water unit. At the same time, check the
hose line.
3. Open the lock of the tank and fill the cleaning agent. The cleaning
agent tank will have a capacity of 1 liter.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509