5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
218
If any instruments are removed from the water unit or new
instruments are connected to the dental treatment unit while the latter
is switched off, they must be plugged into the adapters of the water
unit or returned to the instrument holder again before switching the
dental treatment unit back on.
● Deposit instruments
The AutoPurge function is completed.
➢ Remove the instruments from the water unit and place them back in
their holders.
 When all of the instruments have been deposited, the
instruments
display disappears.
 The AutoPurge procedure is finished. The dental treatment unit is
again ready for operation and can be prepared for the working day.
If individual water-carrying instruments are not flushed, this is displayed
on the touchscreen by a warning triangle.
Back
In this case the
key must be activated to end the AutoPurge
function.
Canceling the AutoPurge function
Deposit instruments
If the error message
rate check or during purging, the AutoPurge function can be canceled.
Stop
➢ Touch the
key on the touchscreen.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
Deposit
appears at the end of the flow
65 48 510 D3509
D3509.201.03.03.23 10.2016