Opening Start sub-screen
Start program
The
is displayed on the touchscreen.
Sub-screen
➢ Touch the
Start
 The
sub-screen is displayed.
Changing waterline treatment system to stand-alone water supply
Aqua
If the
key is displayed gray in the
treatment unit obtains water from the public drinking water supply. If the
key is highlighted orange, the dental treatment unit is switched to stand-
alone water supply. To change the mode of operation, proceed as
follows:
Aqua
1. Press and hold the
Aqua
 The
key flashes orange until the tank for waterline
treatment solution is emptied. The
water supply
display appears on the left-hand side.
 If there is still waterline treatment solution in the tank, it is rinsed
down the drain. This can take up to 6 minutes. The
display then appears on the touchscreen.
2. Mix distilled water with the DENTOSEPT P waterline treatment
solution at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the tank for waterline treatment solution of the water
unit. The tank has a capacity of approx. 1.3 liters. It is full when the
water surface is visible on the filter of the filling funnel.
key.
Start
sub-screen, the dental
key (> 2 s).
Changing to stand-alone
4 Operation
4.7 Water unit
Refill water
161