Setting the spray amount
4.5.8.8
The spray amount is preset at the factory. However, it can be adjusted
using the regulating screw underneath the dentist element. This setting is
then valid for all burr drives.
1. Loosen the ring A counterclockwise.
2. Set the spray amount by turning the screw B.
3. Check the set spray amount with a burr drive and correct the setting
if necessary.
4. Tighten ring A again.
4.5.8.9

Preparing for use of NaCl saline solution

The peristaltic pump prepares a sterile saline solution instead of spray
water for cooling.
The peristaltic pump hose is a disposable item. To reorder it, see "Spare
parts and consumables" [ → 266].
The drive of the peristaltic pump and an NaCl bottle holder are
attached to the dentist element. Please contact your local distributor
if necessary.
B
A new peristaltic pump hose is available.
1. Hang the NaCl bottle (max. 1 liter) onto the bottle holder of the dentist
element.
A
2. Undo the locking mechanism A and fold it down. Swivel the latch
fastener B upward.
3. Open the two hose holders C upward (at the top) and downward. Lay
the silicone hose D without prestress, i.e. with the thickened part
wrapped around the pump wheel. See the adjacent drawing. Close
the hose holders C.
IMPORTANT
Direction of flow of the peristaltic pump
The shorter end of the hose with the cannula must be located at the top
of the pump; the longer end of the hose leading to the handpiece must
be located at the bottom. See the adjacent drawing.