General Handling Instructions - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

5 Care, cleaning and maintenance by the practice team
5.1 Basics
30 s
48 h – 20 °C / 68 °F
24 h – 35 °C / 95 °F
202
4. Immerse the tester in the filled container for 30 seconds.
 The cardboard disk with the culture medium will now absorb 1 ml
of the water sample.
5. Remove the tester from the container. Shake out any excess water.
Empty the tank.
6. Place the tester in the container for incubation either for two days at
a room temperature of 20°C/68°F or for 24 hours at a temperature of
35°C/95°F.
7. Count all germs found on the surface of the tester.
If the number of germs significantly exceeds 100, then sanitation is
Link to sanitizing the dental treatment unit, TENEO
required; see "Sanitizing the dental treatment unit" [ → 248].
5.1.4

General handling instructions

The general handling instructions mainly apply to the dental treatment
unit provided there are no other product-specific handling instructions in
this operating manual. The manufacturer's instructions on the disinfecting
agents should be followed (temperature, concentration, application times,
etc.).
Manual cleaning
Manual cleaning can be done with a cloth or soft brush. Unless specified
otherwise, use lukewarm drinking water for cleaning surface dirt.
Machine cleaning
Thermal disinfection of up to 93 °C according to ISO 15883-1 is possible
for components marked accordingly. Use an alkaline cleaning agent for
this purpose. A laboratory washing machine, such as the Miele G7781,
may be used as a cleaning device. The setting of the device as well as
the type and amount of cleaning agent can be found in the operating
manual of the respective machine.
Manual disinfection
The dental treatment unit can be disinfected by wiping. Use a soft,
colorless cloth for this as well as an approved disinfectant. Unless
specified otherwise in the operating manual, other disinfection methods
(spray disinfection, immersion, etc.) may not be used. Please observe the
information regarding the use of disinfectants in the dental treatment unit
operating manual (temperature, concentration, application times, etc.).
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016