4 Operation
4.5 Dentist element
96
Standardizing the measuring system
Before starting the apex measurement, a functional check or
standardization of the measuring system can be performed by shorting
the electrodes. This cancels out any inaccuracies caused by skips in
impedance in the measurement setup.
➢ Short the electrical measurement system. Plug in the files and hold
them directly against the mucosal electrode.
 If a short signal is heard and the distance display appears with no
bars, the standardization was successful. If not, check the
electrical connections for signs of damage.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016