Advertisement

4 Operation
4.4 Patient chair
66
4.4.7

Moving the patient chair via chair programs

The chair programs can be selected via the touchscreen or via the control
Chair programs 0, S, 1, and 2 on the assistant element, references to "Creating chair programs" and "Linking to dentist element movement"
keys of the assistant element. The entry/exit and mouth rinsing positions
can also be selected via the foot control.
You can individually re-program the factory preset chair programs to suit
your own requirements, see "Creating chair programs and shock
positioning" [ → 72].
You can select the chair program during which the dentist element is to
approach the dental treatment unit; see "Linking dentist element
movement to chair programs" [ → 192].
IMPORTANT
Cuspidor, motor-driven, starting position
The cuspidor bowl automatically swivels back to its starting position
The cuspidor bowl automatically returns beforehand to ensure that the
patient does not collide with it during chair movements. This return travel
is dependent on the chair movement and is executed only if a collision
hazard exists.
4.4.7.1

Moving the patient chair to the entry/exit position

The following functions are triggered for simple patient entry and exit in
the entry/exit position:
● The patient chair moves to an upright position
Entry/exit position
● The dentist element moves to the foot end
● The operating light switches off
● The cuspidor bowl swivels out
The tumbler heater can be set so that it automatically switches off when
the entry/exit position (0) chair program is activated; see "Linking the
tumbler heater to the entry/exit position" [ → 191].
Dentist element, hoses, entry/exit
WARNING
The patient's feet may get caught in the instrument hoses of the dentist
element when he enters or leaves the patient chair.
The patient may trip or fall.
➢ Turn the dentist element outward before the patient enters or leaves
it.
Via touchscreen
Start dialog
The
➢ Touch the
is displayed on the touchscreen.
0
key briefly (< 2 s).
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016