On-Site Installation; Installation Of The Dental Treatment Unit; Media Quality; Connection To The Public Drinking Water System - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
2.3

On-site installation

The on-site installation must have been performed according to our
requirements. The details are described in the document "Installation
Requirements".
2.4

Installation of the dental treatment unit

Installation must be carried out by authorized personnel according to the
installation instructions.
2.5

Media quality

The air and water supplies must meet the requirements specified in the
installation requirements.
To ensure compliance with the medical and national legal requirements
waterline treatment system, standard equipment
for water from dental treatment units, the dental treatment unit must be
equipped with a waterline treatment system.
As the owner of the dental treatment unit, you are generally responsible
for the water quality.
The bacterial count must comply with the national regulations for drinking
water and must not exceed 500 CFU/ml under any circumstances (CFU:
colony forming unit).
In the event of an increased colony count, the installation must be
checked and the cause of contamination eliminated. Alternatively, a
stand-alone water supply can also be installed. the empty tank for
waterline treatment solution can be used as a water container.
Before installation of the dental treatment unit, an acceptable
microbiological water quality for the domestic water supply must be
ensured and documented in the form of a colony count. Sampling and the
colony count must be performed by a competent laboratory.
Test the water quality from the dental treatment unit at regular intervals
and after it has not been used for > 1 week, see Microbiological water test
[ → 201]. Please contact your specialized dealer or your relevant dental
association for the respective national requirements and measures.
Treatment water, immunosuppressed patients, with sterile solution
WARNING
Highly immunosuppressed patients or patients with specific pulmonary
diseases should not come in contact with the water of the dental
treatment unit . The use of sterile solutions is recommended.
2.6

Connection to the public drinking water system

The dental treatment unit complies with the requirements of EN 1717
waterline treatment system, standard
(free discharge with an air gap ≥ 20 mm) and the DVGW German Gas
and Water Association). It is intrinsically safe according to the W540
worksheet and thus also complies with W270 and KTW (Guideline on the
use of plastics in contact with drinking water). It can be connected directly
to the public drinking water system.
2 Safety information
2.3 On-site installation
17