Thermodisinfecting suction hoses

The dental treatment unit is not equipped with thermally disinfectable
suction hoses as standard. Thermally disinfectable suction hoses are
available as special accessories. They are marked with a turquoise ring.
Before thermally disinfecting the suction hoses, you must remove them
from the assistant element and clean the filter inserts; see "Cleaning the
suction hoses" [ → 229].
The suction hoses are removed and cleaned.
1. Pull the handpiece off the spray aspirator hose.
2. Lubricate the O-rings A. For lubricants see "Care, cleaning, and
disinfecting agents" [ → 201].
3. Plug the ends of the handpiece suction hoses onto the hose holders.
Use the adapter B for the saliva ejector hose.
4. Plug the hose holders onto the rails in the (Miele) thermodisinfector
and place the suction hoses on the wire basket provided for that
purpose.
5. Thermally disinfect the suction hoses, filter housings and collectors at
max. 93°C.
To reorder the hose holders for the Miele thermodisinfector, see "Spare
parts and consumables" [ → 266].
5 Care, cleaning and maintenance by the practice team
5.4 Vacuum system
231