Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
4.5.12

SPRAYVIT M multifunctional syringe

The multifunctional syringe is used to clean the treatment areas and blow
them dry. It supplies air and preheated water media. The heating
cartridge for the water is located in the handpiece.
IMPORTANT
Multifunctional syringe, observe the operating instructions
Also observe the operating instructions of the SPRAYVIT M.

Safety instructions

4.5.12.1
The SPRAYVIT M is equipped with extensive safety monitoring functions.
However, please observe the following information.
Multifunctional syringe, changing the hose
CAUTION
After changing the SPRAYVIT M hose, no cooling water for the
SPRAYVIT M heating cartridge flows until the hose is completely filled.
The patient may be scalded by the emission of hot steam. The heating
cartridge can overheat and be destroyed.
➢ After changing the hose on the SPRAYVIT M multifunctional
syringe, press the
ample supply of water in the hose before treating the patient.
IMPORTANT
Multifunctional syringe, switch-on delay
Heating cartridge switch-on delay
To minimize the risk of scalding, the SPRAYVIT M water heater is not
activated for several seconds after the dental treatment unit is switched
on and after a hose change when the
Multifunctional syringe, flow rate
CAUTION
If the flow rate is insufficient, hot water may be emitted by the
SPRAYVIT M.
The patient could thus be scalded.
➢ Check the water flow rate prior to use.
➢ Check the flow rate at least once a month and whenever you
suspect that it may be insufficient as described in the section
"Checking the flow rate of the SPRAYVIT multifunctional syringe"
[ → 221]. Clean the nozzle according to the SPRAYVIT M operating
instructions.
IMPORTANT
Multifunctional syringe, electronic flow monitoring
Electronic flow monitoring
If the electronic flow monitoring system detects low flow, the water
heater will be deactivated and the system displays the corresponding
error message, see also "Error messages" [ → 263].
Water
key briefly and repeat until there is an
Water
key is initially pressed.
4 Operation
4.5 Dentist element
113