Operating Instructions (USA) TENEO
Hose coding
BL motor
65 48 510 D3509
D3509.201.03.03.23 10.2016
4.5.10

Electric motor

IMPORTANT
Electric motor, observe the Operating Instructions
Also observe the operating instructions for the different motors.
4.5.10.1

Motor and coupling versions

Different brushless motors are available depending on the type of use
and the coupling of the handpiece.
Brushless motors are designed as three-phase motors (without carbon
brushes). They feature precise controllability and longevity. The motors
can be sterilized.
The motors feature a special instrument hose with electrical coding. The
dental treatment unit uses these codes to detect which motor is
connected and configures the control accordingly. The confusion of
different variations is ruled out due to the mechanical coding at the hose
and motor. Additionally, the motor couplings are color-coded at the motor
and hose.
The blue coded instrument hose (BU) must be used for the BL motor.
BU
The BL motor is designed for direct operation of T1 CLASSIC
handpieces. For example, in order to use the handpieces of the T1 LINE,
either the basic adapter (apex measurement, no spray) or ISO adapter
(no apex measurement, spray) have to be used as a connecting piece.
Their speed range lies between 90 and 40,000 rpm (revolutions per
minute).
For information on performing apex measurement using the instrument,
Information about apex measurement
refer to the section "Preparing to use the ApexLocator" [ → 94].
BL ISO C motor
4 Operation
4.5 Dentist element
103