4 Operation
4.6 Assistant element
B
150
Saliva ejector
A curved tip that can be placed in the corner of the mouth is provided for
saliva ejection.
Surgical suction device
The surgical cannula can be inserted directly into the saliva ejector
handpiece.
Optionally controllable spray aspirator and saliva ejector
The suction flow from the spray aspirator can be stopped and regulated
using the slide B.
A
The suction flow from the spray aspirator can be stopped and regulated
using the C rotating knob.
Switching off the suction flow using the slide or the 4-way foot switch
Suction handpieces that can be regulated can be closed using the slide
or rotating knob.
If the slide or rotating knob is closed, fluid can flow back into the mouth
due to the lack of suction flow.
➢ When using the spray aspirator with slide, interruption of the suction
flow with the 4-way foot switch must not be configured.
➢ It must be ensured that there is a suction flow before holding the
aspirator in the mouth.
➢ Always remove the aspirator from the mouth before closing the slide
or rotating knob.
➢ Always place the aspirator open in the holder.
CAUTION
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016