5 Care, cleaning and maintenance by the practice team
5.7 Sanitation
max.
256
5.7.3

Sanitizing the dental treatment unit manually

IMPORTANT
The procedure described below regarding manual sanitation is only
feasible if the dental treatment unit is operated with a self-sufficient
water supply.
If the waterline treatment system is set for operation with public drinking
water supply, the sanitization must be guided via the
box, see "Interactively guided sanitation of the dental treatment unit"
[ → 248].
When operating with a self-sufficient water supply, sanitation is to be
carried out manually:
● if the self-sufficient water supply takes more than 28 days, in
exceptional cases
● After longer periods of disuse (> one week)
● If the germination index exceeds 100 germs per milliliter, see
"Microbiological water test" [ → 201].
Manual sanitation is not documented in the sanitation report, see "Display
sanitation report" [ → 255].
Sanitation takes at least 24 hours.
NOTICE
Additional devices, sanitation with waterline treatment solution
Additional devices connected to the external device connection must
not be sanitized with the dental treatment unit.
Additional devices can be damaged. Residues of DENTOSEPT P
waterline treatment solution can remain in the additional devices.
➢ Disconnect any additional devices from the dental treatment unit
prior to sanitation.
Preparation
The following preparations should be made prior to beginning sanitation.
1. If your dental treatment unit is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water lines.
2. Set all burr drives and the SIROSONIC TL scaler to the maximum
water flow rate.
3. Put all instruments and suction hoses in place.
4. Do not remove the tumbler holder from the cuspidor. Put an empty
glass with a volume ≥ 200 ml under the tumbler outlet to prevent
discoloration from the DENTOSEPT P waterline treatment solution.