5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
214
AutoPurge
 The
screen is displayed on the touchscreen.
Setting the instrument purge time and starting the AutoPurge function
The purging time of the instruments may be set between 60 and 180
seconds.
AutoPurge
The
screen is displayed on the touchscreen.
1. Use the – and + keys to set the purging time.
Start
2. Touch the
key.
Error message: Deposit instruments
Deposit instruments
If
is displayed after the AutoPurge function has been
started, the dental treatment unit has detected that not all of the
instruments have been placed in their holders.
➢ Check the seating of the instruments in the holders marked with a
warning triangle on the touchscreen.
 The AutoPurge program will continue when all instruments have
been deposited in the holder.
Inserting water-carrying instruments into the receptacle on the water
unit
The water unit has integrated receptacles for water-carrying instruments
and suction hoses. These enable all instruments to be purged with water.
For this purpose, water-carrying instruments must be inserted into
receptacles on the water unit.
Insert instruments in water unit
The
touchscreen.
1. Remove the SPRAYVIT sleeves from the valve bodies, the straight
and contra-angle handpieces from the water-carrying instruments,
and the suction tips from the suction hoses.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
display appears on the
65 48 510 D3509
D3509.201.03.03.23 10.2016