4.7 Water unit
160
4.7.4

Self-sufficient water supply

If the water quality of the public drinking water supply is not suitable for
the treatment of patients, a dental treatment unit can be operated with a
self-sufficient water supply.
In this case, food-grade water is mixed with DENTOSEPT P waterline
treatment solution in a 100:1 ratio (1 liter of water, 10 ml DENTOSEPT P)
and filled into the tank for waterline treatment solution of the water unit.
The waterline treatment solution will reduce the microbial growth in the
water lines.
The sufficient water supply should be only set as temporary operating
state if public drinking water supply is contaminated and not as
permanent operating state.
The cuspidor is still flushed using the public drinking water supply.
Water lines, microorganisms, self-sufficient water supply
CAUTION
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one's
health.
➢ Mix fresh water for the stand-alone water supply every day. At the
end of the work day, the remaining water must be flushed out of the
tank for waterline treatment solution using the tumbler filling
function.
➢ Check the amount of germs in water of the dental treatment unit at
regular intervals. In particular, if no waterline treatment solution is
used for the water. See "Microbiological water test" [ → 201]
Manually sanitize water supply, self-sufficient, dental treatment unit
CAUTION
If the self-sufficient water supply takes more than 28 days, in
exceptional cases, the dental treatment unit must be manually sanitized.
Please follow the sanitization procedure for operation with self-sufficient
water supply described in the section "Sanitizing the dental treatment
unit manually" [ → 256].
Ensure that the
the touchscreen status bar when operating with a self-sufficient water
supply.
NOTICE
Water supply, self-sufficient, no drinking water
Always fill food-grade distilled water (< 100 colony forming units/ml),
never drinking water, e.g. from beverage bottles, due to the minerals it
contains.
Aqua purificata or aqua destillata can be used as distilled water.
The desired mode of operation is set using the touchscreen.
Start program
Operating Instructions (USA) TENEO
"Days to next sanitization"
Sirona Dental Systems GmbH
display does not appear on
65 48 510 D3509
D3509.201.03.03.23 10.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Teneo

Table of Contents