4 Operation
4.15 Configuration of the dental treatment unit (setup)
194
NOTICE
The CDS 60 station may only be operated with a cleaning agent
approved by Dürr Dental and Sirona, e.g., Orotol plus.
IMPORTANT
Also observe the installation and operating instructions for the CDS 60
from Dürr Dental.
4.15.4.16
Linking the spray aspirator to the 4-way foot switch
A setting can be made to enable interruption and/or reactivation of the
Functionality with the saliva ejector and the surgical suction option
suction flow of the spray aspirator by pressing the 4-way foot switch in any
direction. This function cannot be used on the saliva ejector or on the
surgical suction device. Also observe the safety information, see "Suction
handpieces" [ → 149].
Do not interrupt the suction flow of the spray aspirator with slide using the 4-way foot control
CAUTION
When using the controllabe spray aspirator with slide, interruption of the
suction flow with the 4-way foot switch must not be configured.
Due to the lack of suction flow, fluid may run back out of the spray
aspirator and into the patient's mouth.
➢ Touch the
 If the key is highlighted orange, the removed spray aspirator can
be switched on/off with the 4-way foot switch.
If you deposit the spray aspirator in its holder while the suction flow is
interrupted, the suction flow is automatically restarted when you pick it up
again.
4.15.4.17
Linking the headrest tilt to the 4-way foot switch
If a motor-driven headrest is used, control of the tilt ErgoMotion function
using the 4-way foot switch can be replaced by control of the headrest tilt
function.
➢ Touch the
 If the key is highlighted orange, the tilt of the motor-driven
headrest can be adjusted to the left and right by actuating the 4-
way foot switch.
4.15.4.18
Switching on/off the foot control function for intraoral camera focus
It is possible to enable the foot control to focus the SiroCam AF+ intraoral
camera.
If this function is switched off, the display switches between a still or live
image when the foot pedal is pressed. The knob on the camera is used
to focus the image.
If the function is switched on, the camera image is focused when the foot
pedal is actuated. The display only switches between a still or live image
when the foot pedal is pressed down fully. The knob on the camera can
still be used to focus the image.
Spray aspirator
key.
headrest tilt
key.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016