5 Care, cleaning and maintenance by the practice team
2. If the adapters are not yet located in the receptacles of the water unit,
insert them into the receptacles until they lock into place. The
SPRAYVIT adapters can be inserted only into the two receptacles on
the right side of the dentist element with the guide rib facing upward.
Similarly, a SPRAYVIT adapter must be inserted into the assistant
element side. The adapters always remain in the water unit.
IMPORTANT
Arrangement of adapters
The adapters for the instrument couplings are color coded:
Yellow = SPRAYVIT M, water on right button
Orange = SPRAYVIT M, water on left button
White = high-speed handpiece
Green = BL motor
Blue = BL ISO C motor (ISO interface)
Red = SIROSONIC TL scaler
3. Dentist element side: Insert the couplings of all water-carrying
treatment instruments into the adapter in the water unit (for
SPRAYVIT M: Valve lever in the up position, locking knob in the down
position).
4. Assistant element side: Insert the valve bodies of the SPRAYVIT M
into the adapter in the water unit. Set maximum suction flow on the
suction handpieces and then attach to the connections.
5.7 Sanitation