Saving Instrument Settings - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
SaveMode
DropMode
65 48 510 D3509
D3509.201.03.03.23 10.2016
4.5.6

Saving instrument settings

With static quick setting keys
You can determine whether the
instrument programs with the quick setting keys, see "Preselecting the
quick setting key or function level types" [ → 196].
● SaveMode – The
programs:
The settings made in the Instrument program will be saved after the
instrument is placed in its holder only if the
and held beforehand (> 2 s).
After an instrument is removed, the previously saved settings are preset
again.
An instrument is removed from its holder.
The instrument program is displayed with static quick setting keys
Memory
and the
The desired settings are made; see also "General instrument
functions" [ → 88].
EasyMode Start program
1. Only in the
screen
key.
2. Press and hold down the
 An acoustic signal sounds. The settings in the Instrument
program and its sub-screen are saved.
● DropMode –
When the instrument is deposited, the settings made in the
Instrument program will automatically be saved.
The DropMode is not available in the
mode.
With programmable quick setting keys
With programmable quick setting keys, the set speed and intensity values
can be saved to and called up from one of the quick setting keys. The
settings for selecting and activating the coolant are saved at the same
time.
An instrument is removed from its holder.
The instrument program is displayed on the touchscreen with quick
setting keys.
The coolant is preselected and, if relevant, activated, see
"Preselecting coolant" [ → 88] and "Switching preselected coolant on/
off" [ → 89].
1. Keep the middle quick setting key
setting.
 The programming mode is activated. The speed flashes on the
touchscreen and a flashing bar appears behind the quick setting
keys.
2. Press one of the quick setting keys. This must be done within 5
seconds otherwise the programming mode is closed.
Memory
key should be displayed in the
Memory
key is displayed in the Instrument
key.
operating mode: Touch the
Memory
key.
Memory
key hidden in Instrument programs:
EasyMode Start program
Set
4 Operation
4.5 Dentist element
Memory
key was pressed
Sub-
operating
held down (> 2 s) to save the
85