4 Operation
4.3 Foot control
S
0
52
4.3

Foot control

The dental treatment unit can be operated using a wireless foot control or
a foot control with a cable connection.
4.3.1

Wireless foot control

Technical data of the wireless module, see Foot control radio interface"
[ → 26].
4.3.1.1
Setting the wireless foot control on the dental treatment unit
The wireless foot control must be assigned to the dental treatment unit via
Setting the wireless foot control on the dental treatment unit
a registration. This prevents malfunctions caused by neighboring wireless
foot controls.
The dental treatment unit and wireless foot control are ready for
operation.
All instruments are in place.
1. Simultaneously press and hold the left and right buttons of the foot
control (> 2 s).
 An acoustic signal sounds. The following message appears on
the touch screen:
 If no key is touched, the screen will automatically close after a
specific time. The wireless foot control will not be registered.
2. Confirm that this wireless foot control is to be used on the dental
treatment unit by pressing the
can be interrupted with the
 The message is hidden. The wireless foot control is assigned to
the dental treatment unit.

Battery voltage message

4.3.1.2
The wireless foot control is powered by a battery. An almost empty battery
is detected by the system and displayed in the status bar. In this case, the
battery should be replaced within a week.
When the battery is completely empty, an error code is output, see "Error
messages" [ → 263]. The symbol of the wireless foot control is now
displayed flashing. Replace the battery as soon as possible to prevent
system failure.
The battery can be changed by the user, see "Changing the battery of the
wireless foot control" [ → 245].

Sirona Dental Systems GmbH

Operating Instructions (USA) TENEO
OK
button. The registration process
Esc
key.
D3509.201.03.03.23 10.2016
65 48 510 D3509