5 Care, cleaning and maintenance by the practice team
5.7 Sanitation
248
5.7

Sanitation

Sanitation effectively combats the reproduction of microorganisms in the
water lines.
If the waterline treatment system is set for operation with public drinking
water supply, the sanitization must be guided using the sanitization
screen, see "Interactively guided sanitation of the dental treatment unit"
[ → 248].
If the equipment is being operated with a self-sufficient water supply,
sanitation can only be carried out manually, see "Sanitizing the dental
treatment unit manually" [ → 256]. The
in this operating mode.
For more information, please refer to the section entitled "Self-sufficient
water supply" [ → 160].
5.7.1

Interactively guided sanitation of the dental treatment unit

When operating with public drinking water supply (self-sufficient water
supply switched off), sanitization with interactive guidance should be
carried out:
● Regularly every 4 weeks
The display
the touchscreen. It first appears three days before the date of
sanitization (3d = 3 days until sanitization).
● After longer periods of disuse (> one week)
● if the germination index exceeds 100 germs per milliliter; see
"Microbiological water test" [ → 201].
● After changing from stand-alone water supply to public drinking water
supply and DENTOSEPT P waterline treatment solution
Sanitation with interactive guidance comprises several phases and takes
at least 24 hours.
NOTICE
Additional devices, sanitation with waterline treatment solution
Additional devices connected to the external device connection must
not be sanitized with the dental treatment unit.
Additional devices can be damaged. Residues of DENTOSEPT P
waterline treatment solution can remain in the additional devices.
➢ Disconnect any additional devices from the dental treatment unit
prior to sanitation.
Preparation
The following preparations should be made prior to beginning sanitation.
1. If your dental treatment unit is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water lines.
Operating Instructions (USA) TENEO
Sanitation
Days until next sanitization
Sirona Dental Systems GmbH
program is not available
appears in the status bar of
65 48 510 D3509
D3509.201.03.03.23 10.2016