5 Care, cleaning and maintenance by the practice team
5.7 Sanitation
260
Cleaning the suction hoses
After the manual sanitation, all suction hoses should also be cleaned.
● If the dental treatment unit is equipped with the chemical suction hose
cleaning option, see "Cleaning the suction hoses" [ → 223].
● If the dental treatment unit is not equipped with the chemical suction
hose cleaning option, see "Cleaning vacuum system using cleaning
adapter in the cuspidor or via external container" [ → 225].
5.7.4

Biofilm removal by the service technician

If the microbiological test of the water from the dental treatment unit does
not correspond with the hygiene requirements in spite of regular
sanitization of the water lines and/or regular purge/autopurge cycles, the
biofilm must be removed using special chemicals.
Biofilm removal should be performed if the bacterial count is significantly
above 100 colony forming units per mililiter.
Biofilm removal may only be done by a trained service technician. In this
case, please contact your dental depot.
Before initiating biofilm removal it must be ensured that the reason for the
raised bacterial count is not due to inflowing water.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016