Cleaning the suction system via external container

5.4.2.2
If the dental treatment unit is not equipped with a cuspidor, the suction
system must be cleaned via external container.
1. Prepare 1 liter of cleaning solution in a separate container according
to the manufacturer's instructions and mix it thoroughly.
2. Fill the cleaning solution into a suitable container.
3. If the container is equipped with suitable adapters for the suction
hoses, take the suction cannulae from the suction hoses. Otherwise,
the cleaning solution is to be aspirated with plugged on suction
cannulas.
4. Set maximum suction flow on the suction handpieces.
5. Remove the suction hoses from their holders and aspirate the
cleaning solution simultaneously with all the suction hoses from the
container.
6. Allow the cleaning solution to act. Observe the reaction time specified
for the cleaning solution by the manufacturer.
7. Following the cleaning process, pour at least 1 liter of water into the
container. Aspirate the water in the same way to ensure that no
cleaning agent residues remain in the suction hoses.
8. After finishing, place the suction hoses back in their holders.
5 Care, cleaning and maintenance by the practice team
5.4 Vacuum system
227