5.1.2

Care, cleaning, and disinfecting agents

NOTICE
Approved care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona!
For a continuously updated list of approved agents, please visit
Care, cleaning, and disinfecting agents, with REF number
"www.sirona.com"
documentation, follow the
menu items in the navigation bar. The portal can also be accessed
directly via the following address http://www.sirona.com/manuals. Click
on the menu item
cleaning and disinfection agents"
If you do not have any access to the Internet, please contact your dental
depot to order the list (REF 59 70 905).
5.1.3

Microbiological water test

Perform the microbiological test of the water from the dental treatment
unit at regular intervals and after longer periods of disuse > 1 week; see
"Media quality" [ → 17]. Start the checkups in intervals of no more than two
weeks and adjust the time intervals depending on the results. In addition
to running laboratory tests, you can also use the "Total Count Tester" as
a simple means of performing this test.
To order the motor total count tester, see "Spare parts and consumables"
[ → 266].
IMPORTANT
Total Count Tester, shelf life
Shelf life of the total count tester
The maximum shelf life of the total count tester is 1 year after the date
of receipt.
The cardboard disk contains a dehydrated culture medium. It is activated
by the sample and serves as culture medium for a number of bacteria.
The bacterial count provides information on the hygiene quality of the
water.
Be careful not to touch the inside of the test container or the part of the
tester to be immersed (with nutrient medium) prior to incubation.
1. Allow water to run out of the SPRAYVIT M multifunctional syringe into
the cuspidor bowl for about 1 minute.
2. Use the SPRAYVIT M multifunctional syringe to fill cold water into the
test container to its upper mark.
3. To neutralize the disinfecting agent of the water sample, add approx.
1.5 g of fixing salt (sodium thiosulfate). Fixing salt can be obtained in
pharmacies or from chemical dealers.
5 Care, cleaning and maintenance by the practice team
. To access the online portal for technical
"SERVICE"
"Technical Documentation"
/
"General documents"
and then open the
document.
5.1 Basics
"Care,
201

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Teneo

Table of Contents