Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
4.14

Operation with a PC

The dental treatment unit can be connected to a PC via an Ethernet cable.
The SIVISION and/or SIUCOM plus PC application enables
communication between the dental treatment unit and the PC. SIUCOM
plus is required for PC control. The PC can therefore be directly operated
using the touchscreen and the foot control of the dental treatment unit in
connection with the SiroCam AF+ intraoral camera.
For details, please refer to the "Installation and Configuration of SIUCOM
plus / SIVISION connect" manual.
IT network
CAUTION
Integrating the TENEO into an IT network that includes other devices
can lead to previously unknown risks.
The following changes to the IT network can lead to new risks:
• Changes to the IT network configuration
• Connecting additional elements to the IT network
• Removing elements from the IT network
• Updating devices that are connected to the IT network
• Upgrading devices that are connected to the IT network
The provider of the IT network must determine, analyze, assess, and
control the risks (e.g. IEC 80001-1).
The dental treatment unit can be operated with the following PC versions:
● External PC
The dental treatment unit is connected to an external PC via an
Ethernet cable. PC applications such as SIDEXIS, Microsoft
PowerPoint
from the dental treatment unit.
● Internal PC
The dental treatment unit is operated as a standalone unit without an
external PC. A mini PC is integrated in the base of the chair instead.
It is required to display the camera images using the SI Video PC
application. This version is not intended for image storage or the
control of any additional PC applications.
● Operation without PC
The use of SIVISION is not possible.
The SI Video application is used in dental treatment units with external
PC, on which SIDEXIS is not installed, or with an internal PC (standalone
solution). In these cases, SI Video serves as a video application for the
camera.
IMPORTANT
PC should be hot-plug capable
The PC's HDMI outlet and graphic card should be hot-plug capable.
When the PC is switched on before the dental treatment unit, the
SIVISION monitor may remain black in some PC models.
➢ In this case. switch the dental treatment unit on first, then the PC..
➢ Equip your PC with a hot-plug capable HDMI outlet. Then the
devices can be switched on in any order.
®
, Windows Mediaplayer and SI Video can be controlled