5 Operation
5.5 Main home screen
50
5.5.4.8

Language

Language is only available if "Country Settings" are set for NON-US
else English is preset and fixed for the US (the button is grayed-out).
➢ You have the choice of twelve languages. Select one and confirm
by pressing 'OK'.
Ä Language will be applied after confirmation.
5.5.4.9
Date & time
Format for date: ddmmyy
Format for time (24hours notation): hhmm
➢ Enter date & time and save.
5.5.4.10

Country setting

NOTE
The country setting for the US is pre-set.
WARNING
It's forbidden to change the country setting to Non US if you belong to
US legal regulations. The use of the country setting Non US is not
authorised by the FDA.
For all users except US users:
➢ Change the pre-set country setting to Non-US and confirm the
selection.
Enter the pin code 3 3 3 4 and press 'OK '.
Ä Now you will have access to the full range of pre-set
indications.
5.5.4.11

Battery calibration

After the upgrade the SIROLaser Xtend is equipped with a rechargeable
battery.
For perfect battery performance a battery calibration must be proceeded
in any case of having removed and again mounted or completely
replaced the battery pack, see chapter „Replacing the rechargeable
battery of the control unit"  [→ 67].
1. Switch on the laser without having connected the power supply.
2. Choose "Battery calibration" in the set-up menu.
Ä The following message will appear. "Unplug the laser and press
OK for battery calibration. For further steps refer to the user
manual."
3. Press 'OK'.
Ä The battery will now be discharged automatically until the
device switches off due to lack of power. You will also be
informed that "Battery calibration may take several hours."
4. When the device has been switched off automatically plug in the
power supply and charge the battery for at least 2 hours (best over
night).
Ä Now the battery is calibrated.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018