5 Care, cleaning and maintenance by the practice team
5.5 Components of the water unit
D
E
236
5.5.5

Change the water and air filters

If you notice any changes in media flows, check the water and air filters
for permeability. The filters must be changed, if necessary.
To reorder the filters, see "Spare parts and consumables" [ → 266].
1. Switch the dental treatment unit off at the main switch.
 The water and air supply are switched off.
2. Open the maintenance flap A.
3. Any remaining water flows out of the water filter when it is opened.
You should therefore place an absorbent cloth underneath the filter.
Then unscrew and remove the screw-on cap of the water filter B and/
or the air filter C.
4. Check the filters and replace them if necessary.
5. Reinsert the rubber ring E in the screw-on cap B, C. Then plug on the
filter D as shown in the drawing.
B
 The filter D engages in the screw-on cap B, C.
C
6. Screw the screw-on cap(s) B, C back onto the water unit.