4.15.2

Configuring the EasyTouch user interface

➢ Touch the
 The sub-screen opens.

Switching the key sound on/off

4.15.2.1
The program can be set to issue an acoustic signal that sounds when the
operator touches a key on the touchscreen or a fixed key.
➢ Touch the
 If the key is highlighted orange, the key tone is activated.
4.15.2.2

Calibrating the touchscreen

If the touchscreen is no longer able to precisely locate the position of a
contact, it must be recalibrated.
1. Touch the
 A calibration field is displayed.
2. Touch the small cross on the touchscreen with a blunt pen.
 The cross is displayed at another location on the touchscreen.
3. Repeat this procedure until the cross no longer appears.
4. Touch the empty touchscreen again.
 The touchscreen is now calibrated. The
program is displayed again.
4.15.2.3

Adjusting the touchscreen brightness

The brightness of the touchscreen will decrease after a few years of
operation. This aging effect is normal and can be counterbalanced by
adjusting the brightness. The lighting of the touchscreen has sufficient
reserves for this.
To reduce the aging effect, the touchscreen should not be brighter than
necessary. A warning sign is displayed if the maximum recommended
brightness is exceeded.
➢ Use the – and + keys to set the brightness of the touchscreen.
4.15 Configuration of the dental treatment unit (setup)
EasyTouch user interface
Key tone
key.
touchscreen calibration
key.
4 Operation
key in the setup program.
User interface
setup
187