4 Operation
4.5 Dentist element
90

Preselecting the instrument light operating voltage

4.5.8.6
The original Sirona halogen and LED lamps are usually operated at 3.6
V. Voltages over 3.8 V destroy the lamp. The operating voltage can be
adjusted for lamps made by other manufacturers.
NOTICE
Lamps, restriction of voltage
The operating voltages of different lamps may vary.
Overvoltages can lead to damage.
➢ When changing lamps, make sure that the operating voltage is
properly set for the new lamp.
In the operating modes
Motor
Ultrasonic screen
or
In the operating mode
removed turbine, motor or ultrasonic handpiece are shown.
1. Press and hold the
 The Setup program of the corresponding instrument is displayed.
2. Use the – and + keys to adjust the maximum operating voltage of the
instrument light.
The instrument light voltage of the SPRAYVIT M multifunctional syringe
can also be adjusted, see "Switching the instrument light on/off and
setting the water temperature" [ → 114].
4.5.8.7
Setting the foot control as a direct starter or speed foot control
The foot control can be set as a direct starter or as a speed foot control in
the motor or SIROSONIC TL scaler sub-screen:
● Direct starter
When the foot control is actuated, the instrument is switched on with
the set speed and intensity.
● Speed foot control
Depending on the setting of the foot switch, the instrument controls
the speed and intensity continuously up to the maximum value set.
Sub-screen
The
touchscreen.
➢ Touch the
 If the key is highlighted gray, the direct starter function is switched
on. If the key is highlighted orange, the speed foot control
function is switched on.
Simple/Advanced program
is displayed on the touchscreen.
EasyMode Start program
Setup
fixed key (> 2 s).
of the removed instrument is displayed on the
Direct Starter/speed foot control
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
Turbine,
, the
, the functions of the
key.
65 48 510 D3509