Switching instrument light on/off in the setup program of the
SPRAYVIT M
In order to prevent interference with the treatment process, the
corresponding program does not automatically appear on the
touchscreen when the SPRAYVIT M is removed from its holder. It must
Setup
be opened with the
The SPRAYVIT M multifunctional syringe on the dentist and assistant
element is removed.
Setup
1. Press and hold the
SPRAYVIT
 The
setup program is displayed.
Instrument Light
2. Touch the
 If the key is highlighted orange, the instrument light is turned on
at the multifunctional syringe SPRAYVIT M if this is the only
instrument to have been removed.
Activating/deactivating and setting the water tempering
The heating power of the water heater of the SPRAYVIT M multifunctional
syringe is adjustable. Changes in the inlet temperature of the water
supply (e.g. summer/winter) can thus be compensated for. The setting
range is approx. 8°C.
Multifunctional syringe, heating power
CAUTION
The heating power can be set too high.
The patient then feels that the water is too warm.
➢ Adapt the heating power of the water heating to the inlet
temperature of the water.
➢ Please check the water temperature, e.g., on the back of your hand,
before using the SPRAYVIT M multifunctional syringe.
Water tempering
1. Touch the
 If the key is highlighted orange, the water tempering function is
activated.
2. Use the – and + keys to set the water temperature.
key.
fixed key (> 2 s).
key.Instrument light" />
key.