4 Operation
4.5 Dentist element
110
Apex stop
3. Touch the
 If the key is highlighted orange, the motor stops automatically
when the physiological apex is reached. The – and + keys and
AutoReverse
the
4. Use the – and + keys to set the required apex distance from 1.5 to 0.
If the distance is 0, the motor does not stop until it reaches the
physiological apex. Please note that the distance values are not a
metric length measurement!
 The set distance is displayed to the left of the – and + keys. The
preset motor stop position is displayed in the
black triangle to the right of the distance display under the text
"Stop".
5. To combine the automatic motor stop with the automatic switching to
counterclockwise rotation, press the
 If the button is orange, the motor automatically switches to
counterclockwise rotation following the motor stop the next time
the foot pedal is operated. When the file is withdrawn, the bur
drive automatically switches back to clockwise rotation.
6. To switch on the apex acoustic signals, press the
signals
key.
 If the key is colored orange, an acoustic signal is issued when the
apex or the set motor stop position is reached. If the AutoReverse
function is activated, three acoustic signals are issued when the
motor switches to counterclockwise rotation.
7. To switch on the distance signal tones, press the
signal tones
key.
 If the key is colored orange, distance acoustic signals will be
issued in addition to the distance display diagram. If the
automatic motor stop is switched off, the intervals between the
acoustic signals vary according to the measured distance from
the physiological apex. If this function is switched on, the acoustic
signals vary depending on the measured distance to the preset
motor stop position. For more information on the acoustic signals
during apex measurement, see the section "Acoustic signals"
[ → 100].
Return
8. Touch the
key.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
key.
key are displayed.
AutoReverse
D3509.201.03.03.23 10.2016
Motor dialog
as a
key.
Apex acoustic
Apex distance
65 48 510 D3509

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Teneo

Table of Contents