5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
216
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
If the hoses are pinched, the water flow will be obstructed during
purging.
 Place all water-carrying instruments and suction hoses into the
receptacles.
Start
5. Touch the
key on the touchscreen.
 First, the dental treatment unit checks whether there is water flow
through the instruments. This will take a moment.
Error message: No water flow
If the dental treatment unit detects no water flow through an instrument,
you can try to restore water flow through the relevant instrument. If this is
not possible, the instrument concerned can be excluded from the purging
process.
1. Check the water flow through the instruments in the instrument
positions marked with a warning triangle on the touchscreen. Set the
instruments to maximum water flow. Leave all instruments plugged
into the water unit.
 If the dental treatment unit detects the water flow, the warning
triangle will disappear. If water flow is detected for all instruments,
the AutoPurge program automatically continues.
2. If you want to exclude the instruments concerned from purging, touch
Exclude instrument
the
 The AutoPurge program continues. The water lines affected are
not included in the purging process.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
key.
D3509.201.03.03.23 10.2016