5 Care, cleaning and maintenance by the practice team
5.5 Components of the water unit
234
5.5.3

Cleaning the outlet lines of the cuspidor

With the option chemical suction hose cleaning, the drain outlets of the
cuspidor are not cleaned/disinfected. If the dental treatment unit is
equipped with this option, the drains of the cuspidor must therefore be
cleaned weekly. Use the same agent as for the suction lines.
NOTICE
Care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 201].
NOTICE
Vacuum system, foaming cleaning agents
Domestic cleaning agents foam up.
Foaming cleaning agents often cause foam and water to be sucked into
the dry suction system. This can cause damage to the suction machine.
➢ Use only care, cleaning and disinfecting agents approved by Sirona,
see"Care, cleaning, and disinfecting agents" [ → 201].
1. Prepare 1 liter of cleaning solution in a separate container according
to the manufacturer's instructions and mix it thoroughly.
2. Fill the cuspidor with the cleaning solution and leave it to take effect.
Observe the reaction time specified for the cleaning solution by the
manufacturer.
3. Rinse the cleaning solution away. To do this, pour at least 1 liter of
water into the cuspidor.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016