Cleaning/Disinfecting The Sirocam Af+ Intraoral Camera; Checking The Flow Rate On The Sprayvit M - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
100 ml
max. 45 sec
65 48 510 D3509
D3509.201.03.03.23 10.2016

Cleaning/disinfecting the SiroCam AF+ intraoral camera

5.3.4.4
Due to its shape, the SiroCam AF+ intraoral camera complies with
hygienic requirements and therefore has no spots that are difficult to
reach. It can be wipe disinfected.
NOTICE
Care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 201].
NOTICE
Intraoral camera, lens window
The lens window is sensitive to scratches.
Deep scratches in the lens window impair image quality.
➢ Protect the lens window against scratching. Disinfect it with a soft,
lint-free cloth.
Checking the flow rate on the SPRAYVIT M multifunctional
5.3.4.5
syringe
If the flow rate is less than 135 ml/min with the button fully actuated, there
is the risk of excessively hot water being emitted.
Multifunctional syringe, flow rate during cleaning
CAUTION
If the flow rate is insufficient, hot water may be emitted by the
SPRAYVIT M.
The patient could thus be scalded.
➢ Check the water flow rate prior to use.
➢ Clean the nozzle according to the SPRAYVIT M Operating
Instructions.
Perform the following measurement to exclude a patient risk.
➢ Fill a measuring cup up to the 100 ml mark with the water button fully
actuated while measuring the required filling time.
 The filling time must not exceed 45 seconds.
If the specified water quantity is not reached within the filling time of 45
seconds, clean the nozzle or the dental treatment unit will be checked by
a service engineer.
5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
221