5 Care, cleaning and maintenance by the practice team
5.4 Vacuum system
230
4. Take the collectors C out of the suction hoses. Collect the amalgam
residues in a glass filled with water.
5. Take off the suction needles and flush the suction hoses with clean
water.
If the dental treatment unit is equipped with thermally disinfectable
suction hoses, thermal disinfection can be performed after cleaning, see
"Thermally disinfecting the suction hoses" [ → 231].
Assembly is performed in reverse order. Grease the O-rings A before
reconnecting the suction hoses. Lubricants, see "Care, cleaning, and
disinfecting agents" [ → 201]. Make sure that the suction hoses snap into
place.