4 Operation
4.5 Dentist element
106
Setting intermediate speed values with static quick setting keys
The electric motor is removed from its holder.
Motor program
The
with static quick setting keys or the
Start program
is displayed on the touchscreen.
➢ Increase speed: Hold down a quick setting key, which has a speed
value that is greater than or equal to the one displayed in the first line
(> 1 s).
Decrease speed: Hold down a quick setting key, which has a speed
value that is less than the one displayed in the first line (> 1 s).
 The selected speed is displayed in the first line in rpm
(revolutions per minute). The quick setting keys are shaded gray
for intermediate values.
IMPORTANT
Increments
The size of the increments depends on the speed range setting.
For BL, BL ISO C and BL Implant motors:
From 90 to 400 rpm = 10 rpm increments
From 400 to 5,000 rpm = 200 rpm increments
From 5,000 to 40,000 rpm = 1,000 rpm increments
Please note that the speed of the burr depends on the selected straight
or contra-angle handpiece.
Setting intermediate speed values with programmable quick setting
keys
The electric motor is removed from its holder.
Motor program
The
is displayed on the touchscreen with
programmable quick setting keys.
➢ Press and hold the left or right quick setting key – or + (> 1 s).
 The speed is increased or reduced.
For increments see "Setting intermediate speed values with static quick
setting keys" (above).
Setting the speed with function levels and saving on a level
The electric motor is removed from its holder.
Motor program
The
with function levels is displayed on the
touchscreen.
➢ Set the speed using the – and + keys.
< 1 s coarse adjustment, > 1 s fine adjustment
 The selected speed is displayed in the first line in rpm
(revolutions per minute).
For increments for coarse adjustment, see "Selecting the speed with the
quick setting keys" (above).
For increments for fine adjustment, see "Setting intermediate speed
values with static quick setting keys" (above).