Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
5 Care, cleaning and maintenance by the practice team
Start
 The
sub-dialog is displayed.
Purge function
2. Touch the
Purge
 The
screen is displayed on the touchscreen.
Setting the purge time and starting the purge function
The purge time of the removed instruments can be set between 20 and
180 seconds.
Purge
The
dialog is displayed on the touchscreen.
1. Use the – and + keys to set the purging time.
Start
2. Touch the
key.
Please note that changes made during the purging process take effect
only after the next purge process is started.
Error message: Refill water (only for stand-alone water supply)
Refill water
If the
display appears after the purge function has been
started, there is not sufficient water in the tank for waterline treatment
solution of the water unit to purge the water lines. The purge function
cannot be started with insufficient water.
➢ Mix distilled water with the DENTOSEPT P waterline treatment
solution at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the tank for waterline treatment solution of the water
unit. For more information, please refer to the section "Self-sufficient
water supply" [ → 160].
 When sufficient water has been refilled, the purge program
continues.
5.3 Instruments and instrument hoses
key.
211