5 Care, cleaning and maintenance by the practice team
5.4 Vacuum system
228
5.4.3

Sterilize and disinfect the suction handpieces

Sterilization/disinfection
All parts of the suction handpieces can be sterilized and thermally
disinfected.
A
Suction tip
B
Suction Handpiece
C
Rotary joint
D
Saliva ejector handpiece
E
Surgical suction tip
F
Intermediate piece
Lubricating suction handpieces
The disconnection points of the suction handpieces must be lubricated
once a week and after each thermal disinfection or sterilization process.
Grease
CAUTION
Unsuitable greases
Greases that are not food safe may endanger the patient's health
Rubber materials such as e.g. O-rings are corroded by unsuitable
greases.
➢ Do not under any circumstances use vaseline or silicone greases.
➢ Use only greases approved by Sirona.
1. Pull the handpiece of the spray aspirator, the saliva ejector and, if
installed, the surgical suction device off of their suction hoses.
2. Take apart the handpiece of the spray aspirator at its joints.
3. Regrease the disconnection points and O rings of the handpieces.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016