5 Care, cleaning and maintenance by the practice team
5.2 Surfaces
B
A
93°C
200°F
206
5.2.3

Disinfecting handles

NOTICE
Care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 201].
Dentist element
The handles on the dentist element can be spray, wipe and thermally
disinfected. They can be detached.
1. Press the locking rocker switch A toward the rear with your thumb.
 The handle B is unlocked.
2. Raise the handle B slightly.
3. Remove the handle from the upper guide tab C.
C
4. Repeat this procedure for the opposite handle.
Proceed in reverse order when reattaching the handle. The locking rocker
switch A engages automatically.
Assistant element
The handles on the instrument holder of the assistant element can be
spray, wipe and thermally disinfected. See also "Thermally disinfect the
instrument holder of the assistant element" [ → 209].
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016