5 Care, cleaning and maintenance by the practice team
5.2 Surfaces
208
5.2.6
Thermally disinfecting the instrument holder of the dentist
element and sterilizing the silicone mat
The instrument holder can be removed for easier cleaning or thermal
disinfection.
Removing the instrument holder
1. Remove all instruments from the holder.
2. Swivel both of the levers A below the instrument holder to the rear.
 The latch of the instrument holder is released.
3. Grasp underneath the instrument holder and raise it at the rear.
 The instrument holder tilts to the front and can be removed from
the dentist element in an upward direction.
The removable silicone mat on the dentist element can be sterilized.
Inserting the instrument holder
The levers A are swiveled to the rear.
1. Fit the front recesses B of the instrument holder onto the bolts in the
dentist element.
2. Gently press the instrument holder into the dentist element.
3. Hold the instrument holder firmly and swivel both levers A toward the
front.
 The instrument holder is locked into the dentist element.
Ball stopper
If a ball stopper is inserted in an unused instrument holder, it can simply
be pushed out of the holder from the rear to facilitate cleaning or thermal
disinfection.
Motor holder
A sterilizable motor holder is available for surgical procedures; see also
"Preparing the dental treatment unit for sterile operation" [ → 130].
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016