5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
132°C – 134°C
270°F – 274°F
3 min
220
The silicone isolation sleeve is a disposable item. It must be changed
after each patient. The silicone isolation sleeve must be sterilized prior to
initial use.
132°C – 134°C
270°F – 274°F
To reorder the silicone isolation sleeve, see "Spare parts and
3 min
consumables" [ → 266].
Sterilize the root canal files according to the manufacturer's instructions.
5.3.4.3

Disinfecting/sterilizing Mini L.E.D. curing light

NOTICE
Care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 201].
1. Remove the connection cable of the Mini L.E.D. by turning the
handpiece.
2. Pull out the light guide and remove the glare shield.
3. Sterilize the light guide and the glare shield at 134° C, 2 bar for 3 min.
4. Disinfect the handpiece of the Mini L.E.D.
5. Screw the sterilized light guide and glare shield back onto the
Mini L.E.D.
6. Reconnect the handpiece of the Mini L.E.D. to the connecting cable.
The following points should also be observed when operating the
Mini L.E.D.:
● Always use the glare shield to protect your eyes.
● Check the light guide after each use. Make sure that the light guide is
in perfect condition.
● There should be no traces of composite material on the light guide.
Immediately remove any residue.
● If you find any damage, replace the light guide, since damage will
impair its performance considerably.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016