5 Care, cleaning and maintenance by the practice team
5.7 Sanitation
254
Error message: Inserting instruments in the water unit
If an instrument is accidentally deposited in its holder during the reaction
Insert instruments in water unit
time, the message
touchscreen after the dental treatment unit is switched on. The sanitizing
process is then completed anyway.
1. Take the accidentally removed instrument out of the instrument
holder and place it back in the receptacle on the water unit so that it
can be rinsed in sanitation phase 5.
2. If you want to exclude the instruments from sanitation, touch the
Exclude instrument
key.
 Excluded instruments will not be rinsed in sanitation phase 5.
Sanitation phase 5 – Rinsing water lines with water
The DENTOSEPT P waterline treatment solution is rinsed out of the
SPRAYVIT hoses, instrument hoses, and tumbler filling unit with water.
This takes several minutes.
Sanitization phase 6 – Returning the instruments to their holders
After the water lines have been purged, the instruments and suction
hoses can be removed from the receptacle on the water unit and re-
inserted in the instrument holders.
Return instruments to their holders
The
touchscreen.
➢ Place the SPRAYVIT sleeves, the straight and contra-angle pieces
and the suction tips on the instruments or suction hoses. Return all
instruments and suction hoses in the instrument holders.
 The sanitation process is finished. The dental treatment unit is
again ready for operation.
Displays in case of incomplete sanitation
The sanitation dialog detects whether the sanitation of all instruments and
the tumbler filling was completed. Errors that occurred during sanitation
will be displayed on the touchscreen as follows:
● Crossed-out, solid gray circle: The instrument was excluded from
sanitation before it had been filled with DENTOSEPT P waterline
treatment solution
● Orange circle with warning triangle: The instrument or tumbler filling
was not (sufficiently) rinsed; the water lines still contain waterline
treatment solution
➢ In the latter case, rinse the affected instruments and tumbler filling
manually after sanitation.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
is displayed on the
display appears on the
65 48 510 D3509
D3509.201.03.03.23 10.2016