Sirona teneo Operating Instructions Manual page 139

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
Setting the automatic motor stop of the ApexLocator
If your dental treatment unit is equipped with the ApexLocator option, the
motor can be set to stop automatically at a preset distance from the apex.
The motor stop can be combined with the
"Switching the AutoReverse function on/off".
Apex stop
1. Touch the
 If the key is highlighted orange, the automatic motor stop is
activated. The – and + keys are displayed.
2. Use the – and + keys to set the required apex distance from 1.5 to 0.
If the distance is 0, the motor does not stop until it reaches the
physiological apex. Please note that the distance values are not a
metric length measurement!
 The set distance is displayed to the left of the – and + keys. The
preset motor stop position is displayed in the
Endodontics program
distance display under the text "Stop".
Apex settings are stored automatically when the treatment is completed.
Switching the apex distance signal tones on/off
If your dental treatment unit is equipped with the ApexLocator option, you
can make a setting to determined that distance signal tones are issued in
addition to the graphical distance display. If the automatic motor stop is
switched off, the intervals between the beeps vary according to the
measured distance from the physiological apex. If this function is
switched on, the beeps vary depending on the measured distance to the
preset motor stop position. For more information on the signal tones
during apex measurement, see the section "Beeps" [ → 100].
Apex distance signal tones
➢ Press the
 If the key is highlighted orange, the apex distance acoustic signal
is activated.
Switching the torque signal and apex signal tones on/off
This can be used to set the acoustic signal that sounds whenever 75% of
the currently set torque value is exceeded.
If your dental treatment unit is equipped with the ApexLocator option, this
button also switches the apex signal tones on or off. If the signal tones are
switched on, a beep is issued when the apex or a set motor stop position
is reached. For more information on the acoustic signals during apex
measurement, see the section "Acoustic signals". For more information
on the acoustic signals during apex measurement, see the section
"Acoustic signals" [ → 100].
Torque signal and apex acoustic signals
➢ Press the
 If the key is marked in orange, the torque signal and the apex
acoustic signals are switched on.
AutoReverse
key.
screen as a black triangle to the right of the
key.
4 Operation
4.5 Dentist element
function, see
Simple/Advanced
key.
139

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Teneo

Table of Contents