Technical Data - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
General Technical Data on Mini L.E.D.
Optical specification of Mini L.E.D.
Hazard distance (HD) and exposure
hazard value (EHV) according to IEC
62471
65 48 510 D3509
D3509.201.03.03.23 10.2016

Technical data

4.6.7.6
Model:
Weight of handpiece without hose: 105 g
Dimensions:
Current consumption of handpiece: 5 V DC / 0.65 A
Thermal safety:
Wavelength:
Max. intensity:
Light power dia. 7.5 mm (standard
version):
Light power dia. 5.5 mm (available
from Satelec):
Light power:
Classification acc. to IEC 62471
1
Only composite materials can be cured which react to the stated
wavelength. The Mini L.E.D. is not suitable, e.g. for Lucirin
maximum 380 nm).
Blue light
Free group
Blue light
Risk group 1
Blue light
Risk group 2
Thermal hazard to the
cornea
Free group
Mini L.E.D.
Dia. 23mm x 240mm
Overheat protection
420 nm – 480 nm
at 450 nm
2
1250mW/cm
2000 mW/cm
450 mW – 500 mW
Risk group 2 with light bar removed
HD
EHV
907 mm
20.6
323 mm
2.6
0.01
0.97
4 Operation
4.6 Assistant element
1
2
®
(absorption
155

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Teneo

Table of Contents