4 Operation
4.13 SIVISION digital video system
172
4.13.1

SIVISION monitor

22" monitor
The monitor is equipped with loudspeakers and a glass screen to
facilitate cleaning. Please refer to the operating instructions for the 22" flat
screen monitor 2015 model.
Monitor, not suitable without approval
WARNING
Monitors without approval must not be connected.
They endanger the product safety of the dental treatment unit.
➢ Use only monitors that are approved according to IEC 60950-1,
IEC 62368-1 (office equipment) or IEC 60601-1 (medical devices).
Connection of loudspeakers
WARNING
Unsuitable devices can be connected to the loudspeaker port of the
monitor.
The connection of unsuitable devices endangers the product safety of
the dental treatment unit.
➢ The loudspeaker port of the monitor may be connected only to a
device that complies with IEC 60950-1, IEC 62368-1 (office
equipment such as PCs) or IEC 60601-1 (medical devices). Under
no circumstances should it be connected e.g. to a stereo system.
Attachment versions
Monitor on lamp support tube
The monitor can be rotated and swiveled.
Monitor on tray arm
The monitor can be rotated and swiveled.
Monitor, collision with patient
CAUTION
The patient can collide with the monitor.
The patient's head could collide with the monitor during movement of the
patient chair or during use of the cuspidor.
➢ Make sure that the patient does not collide with the monitor. Swivel
the monitor out of the collision zone as soon as it is no longer
required.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016