Loss of Power Meter Calibration Data

The power meter calibration data will be lost by committing any of the following actions:
Changing sweep type. If the sweep type is changed (linear, log, list, CW, power) while
power meter calibration is on, the calibration data will be lost. However, calibration data is
retained if you change the sweep type while power meter calibration is off.
Changing frequency. Power meter calibration data will also be lost if the frequency is
changed in log or list mode, but it is retained in linear sweep mode.
Pressing m. Presetting the instrument will erase power meter calibration data. If the
instrument state has been saved in a register using the CG) key, you may recall
instrument state and the data will be restored. Saving the instrument state will not protect
the data if the instrument is turned off.

Interpolation in Power Meter Calibration

If the frequency is changed in linear sweep, or the start/stop power is changed in power sweep,
then the calibration data is interpolated for the new range.
If calibration power is changed in any of the sweep types, the values in the power setting array
are increased or decreased to reflect the new power level. Some accuracy is lost when this
occurs.

Power Meter Calibration Modes of Operation

You can set the analyzer to update the correction table at each point for sweep (as in a
leveling application), using the ~~~~~~~I softkey. In this mode, the analyzer checks the
power level at every frequency point each time it sweeps You can also have more than one
description in Chapter 9).
While using the continuous sample mode, the power meter must remain connected as shown
in Figure 6-57. A power splitter or directional coupler samples the actual power going to the
test device and is measured by the power meter. The power meter measurement provides the
information necessary to update the correction table via HP-IB.
Continuous correction slows the sweep speed considerably, especially when low power levels
are being measured by the power meter. It may take up to 10 seconds per point if the power
level is less than -20 dBm. For faster operation, you can use the sample-and-sweep mode. If
you use a directional coupler you must enter the attenuation of the coupled arm with respect
to the through arm using the ~~~~~~~~~~ softkey.
. . . _......... . ; ; _ . , . , . ;
. . . . . . . _. ~ . s +, . , . F z . < : . ( , ,
the

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8720d8722d

Table of Contents