Measuring LOW-LOSS CAPACITOR Calibration Data

The LOW-LOSS CAPACITOR calibration should be performed for high Q (or low D:
dissipation factor) measurements at high frequencies. The LOW-LOSS CAPACITOR
calibration allows high accuracy for phase measurements. This calibration can be skipped
if you do not need it for your purposes.
Step 1. As shown in Figure 3-38, turn the LOW-LOSS CAPACITOR clockwise with the provided
torque wrench to connect it securely to the 7-mm terminal.
CAUTION
Do not turn the nob of the LOW-LOSS CAPACITOR. Lightly hold the nob to prevent the
inner coaxial electrodes from moving when turning the outer connector nut of the
LOW-LOSS CAPATITOR.
Figure 3-38
Connecting LOW-LOSS CAPACITOR
Step 2. Click the
Figure 3-39
Step 2
Step 3. A √ mark appears on the left side of the
LOW-LOSS CAPACITOR calibration data measurement.
Step 4. Turn the LOW-LOSS CAPACITOR counterclockwise to remove it.
Chapter 3
Basic Operations for RF Devices Measurement
button (Figure 3-39).
Meas Low Loss C
STEP 3. Calibration
button upon completion of the
Meas Low Loss C
65