Check 11, DC-DC Converter Open Output Voltages

1. Turn o the 4263B.
2. Disconnect the DC-DC Converter Output Cable shown in Figure 4-15 from connector \J504"
on the A1 Main Board Assembly.
3. Turn the 4263B on.
4. Using the multimeter, check if the DC-DC Converter open output voltages on the output
connector pins shown in Figure 4-15 are in the ranges shown in Table 4-2.
5. If passed, replace the A1 Main Board Assembly, otherwise replace the DC-DC Converter
Assembly.
4-18
T roubleshooting
6
1
2
5.1
6
4
3
12
5
3
12
0.2 V
0.6 V
6
0.6 V

Advertisement