Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page

Turning On/Off the Auxiliary Bin

Functional Description
If the sorting result depends on the secondary parameter, you can specify the
secondary parameter limit values in the 2nd LOW/HIGH field. For sorting
based on the secondary parameter, there are two scenarios:
Sorting with the secondary parameter limit values specified and the
auxiliary bin turned off:
Only those DUTs that fall within the secondary parameter values are sorted
based on the result of comparing the primary parameter values. On the
other hand, DUTs that do not fall within the secondary parameter limit
values are sorted as OUT OF BINS, regardless of whether they fall within the
primary parameter limit values.
Figure 4-18
Auxiliary bin turned off
Sorting with the secondary parameter limit values specified and the
auxiliary bin turned on:
DUTs that do not fall within the primary parameter limit values are sorted as
OUT OF BINS. In addition, DUTs that fall within the primary parameter limits
but are out of the secondary parameter limits are sorted into the auxiliary
(AUX) bin.
Keysight E4980A/AL Precision LCR Meter
137