Editing Sensor Calibration Tables - Keysight Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

To use sensor calibration tables you:
1 Edit a sensor calibration table if necessary.
2 Select the sensor calibration table.
3 Enable the sensor calibration table.
4 Zero and calibrate the power meter. The reference calibration factor used
5 Specify the frequency of the signal you want to measure. The calibration factor
6 Make the measurement.

Editing sensor calibration tables

It is not possible to create any additional sensor calibration tables. However, the
20 existing ones can be edited using the MEMory subsystem. To do this:
1 Select one of the existing tables using:
2 Enter the frequency data using:
3 Enter the calibration factors using:
Keysight EPM-P Series Programming Guide
during the calibration is automatically set by the power meter from the sensor
calibration table.
is automatically set by the power meter from the sensor calibration table.
MEMory:TABle:SELect <string>.
For information on naming sensor calibration tables see
calibration tables"
on page 66. For information on the current names which
you can select refer
MEMory:TABle:FREQuency <numeric_value>
{,<numeric_value>}
MEMory:TABle:GAIN <numeric_value>
{,<numeric_value>}. The first parameter you enter should be the reference
calibration factor, each subsequent parameter is a calibration factor in the
sensor calibration table. This means that entries in the frequency list
correspond as shown with entries in the calibration factor list.
to."Listing sensor calibration table names"
Power Meter Remote Operation
"Naming sensor
1
on page 65
63