MEMory Subsystem

MEMory:CATalog:TABLe?

This command is used to list the stored sensor calibration (8480 Series power
sensors and N8480 Series sensors with Option CFT only) and frequency
dependent offset tables.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are stored tables.
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory, in bytes, used
– The second parameter indicates the memory, in bytes, available for the
– Each string parameter returned indicates the name, type and size of a stored
For example, a sample of a response may look like:
1178,10040,"DEFAULT,TABL,14","HP8481A,TABL,116",
"HP8482A,TABL,74","HP8483A,TABL,62"...........
Predefined sensor calibration table is not applicable for N8480 Series power
NOTE
sensors with Option CFT. Therefore you are required to create a new sensor
calibration table for the sensors when a sensor calibration table is needed.
The power meter is shipped with a set of predefined sensor calibration tables. The
data in these sensor calibration tables is based on statistical averages for a range
of Keysight power sensors. These tables can be altered. The predefined data is
listed in your User's Guide. These power sensors are:
306
for the storage of tables.
storage of tables.
table:
– <string>,<type>,<size>
-- <string> indicates the name of the table.
-- <type> indicates TABL for a table.
-- <size> indicates the size of the table in bytes.
Keysight EPM-P Series Programming Guide